PRIDE MOBILITY PRODUCTS PRIDE MOBILITY PRODUCTS; WHEELCCHAIR, POSITIONING. ELECTRIC Back to Search Results
Model Number PLR3955LT
Medical Device Problem Code Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Codes Abrasion (1689); Laceration(s) (1946)
Date of Event 06/16/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device has not yet been made available for evaluation.Should further information or the device become available, a follow-up report will then be issued.
 
Event or Problem Description
Consumer alleges the chair wasn't reclining, allegedly made a clicking noise and he alleges the footrest wasn't coming up.
 
Additional Manufacturer Narrative
The root cause of the event appears to be lift and recline (blue and green) actuator connectors were reversed.The evaluator matched the colored connectors and the chair functioned as designed.Based on the functional testing recorded in the device history record, the switching of the colored connectors occurred post final release.
 
Event or Problem Description
Consumer alleges the chair wasn't reclining, allegedly made a clicking noise and he alleges the footrest wasn't coming up.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRIDE MOBILITY PRODUCTS
Common Device Name
WHEELCCHAIR, POSITIONING. ELECTRIC
Manufacturer (Section D)
PRIDE MOBILITY PRODUCTS
401 york ave
duryea PA 18642
Manufacturer (Section G)
N/A
n/a
n/a
n/a
Manufacturer Contact
kelly livingston
401 york avenue
duryea 18642
8008008586
MDR Report Key17209664
Report Number2530130-2023-00084
Device Sequence Number12122682
Product Code INO
UDI-Device Identifier00606509401500
UDI-Public00606509401500
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K091578
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 09/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberPLR3955LT
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/20/2023
Supplement Date Received by Manufacturer06/20/2023
Initial Report FDA Received Date06/27/2023
Supplement Report FDA Received Date09/07/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/06/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexMale
-
-