| Model Number |
PLR3955LT |
| Medical Device Problem Code |
Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Codes |
Abrasion (1689); Laceration(s) (1946)
|
| Date of Event |
06/16/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device has not yet been made available for evaluation.Should further information or the device become available, a follow-up report will then be issued.
|
| |
|
Event or Problem Description
|
|
Consumer alleges the chair wasn't reclining, allegedly made a clicking noise and he alleges the footrest wasn't coming up.
|
| |
|
Additional Manufacturer Narrative
|
|
The root cause of the event appears to be lift and recline (blue and green) actuator connectors were reversed.The evaluator matched the colored connectors and the chair functioned as designed.Based on the functional testing recorded in the device history record, the switching of the colored connectors occurred post final release.
|
| |
|
Event or Problem Description
|
|
Consumer alleges the chair wasn't reclining, allegedly made a clicking noise and he alleges the footrest wasn't coming up.
|
| |
|