BECTON DICKINSON MEDICAL SYSTEMS BD SAFETYGLIDE NEEDLE; HYPODERMIC SINGLE LUMEN NEEDLE Back to Search Results
Model Number 306609
Medical Device Problem Code Complete Blockage (1094)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/19/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H.3.A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
 
Event or Problem Description
It was reported while using bd safetyglide needle the needle was blocked.There was no report of patient impact.The following information was provided by the initial reporter: very difficult to push the vaccine through the needle.
 
Additional Manufacturer Narrative
H6: investigation summary no samples or photos received by our quality team for evaluation.Furthermore, a device history record review showed no rejected inspections or quality issues during the production of the provided lot number that could have contributed to the reported defect.During review of the process, process variations during the needle lubricant application can create clogged needles.Corrective and preventative actions have been implemented.Several quality initiatives have been implemented on our manufacturing line to ensure that the needle lubricant application is properly applied during the manufacturing process.Additionally, a monitoring program is also in place to verify the needle lubricant is applied uniformly to the needle.
 
Event or Problem Description
It was reported while using bd safetyglide needle the needle was blocked.There was no report of patient impact.The following information was provided by the initial reporter: very difficult to push the vaccine through the needle.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD SAFETYGLIDE NEEDLE
Common Device Name
HYPODERMIC SINGLE LUMEN NEEDLE
Manufacturer (Section D)
BECTON DICKINSON MEDICAL SYSTEMS
route 7 and grace way
canaan CT 06018
Manufacturer (Section G)
BECTON DICKINSON MEDICAL SYSTEMS
route 7 and grace way
canaan CT 06018
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8448235433
MDR Report Key17211311
Report Number1213809-2023-00668
Device Sequence Number17725590
Product Code FMI
UDI-Device Identifier10612479262567
UDI-Public10612479262567
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
NA
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number306609
Device Catalogue Number306609
Device Lot Number2258318
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/01/2023
Supplement Date Received by Manufacturer07/17/2023
Initial Report FDA Received Date06/27/2023
Supplement Report FDA Received Date08/04/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/15/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-