• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALLERGAN (COSTA RICA) STYLE 68 SALINE FILLED BREAST IMPLANT; PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALLERGAN (COSTA RICA) STYLE 68 SALINE FILLED BREAST IMPLANT; PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE Back to Search Results
Model Number 68LP-275
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 05/20/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: deflation.
 
Event or Problem Description
Healthcare professional reported right side deflation.Device has been explanted and replaced.
 
Additional Manufacturer Narrative
Device evaluation based on the device analysis grid, the assessments of the complaint are: ¿ deflation: observed opening on posterior side assessed as fold flaw opening.Additional observations: ¿ observed creases on the device.¿ observed wear abrasion on the device.
 
Event or Problem Description
Healthcare professional reported right side deflation.Device has been explanted and replaced.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STYLE 68 SALINE FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer (Section G)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer Contact
terry ingram
12331-a riata trace parkway
building 3
austin, TX 78727
8479366324
MDR Report Key17213001
Report Number9617229-2023-10307
Device Sequence Number12345570
Product Code FWM
UDI-Device Identifier10888628000575
UDI-Public10888628000575
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
P990074
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2023
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/27/2023
Device Model Number68LP-275
Device Catalogue Number68LP-275
Device Lot Number3410058
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/05/2023
Supplement Date Received by Manufacturer08/18/2023
Initial Report FDA Received Date06/27/2023
Supplement Report FDA Received Date09/14/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/27/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age80 YR
Patient SexFemale
Patient Weight59 KG
-
-