Event or Problem Description
|
|
It was reported that the cartridges caused damage to several ar40e intraocular lenses (iol's) during the procedure on a single patient.This caused no lasting issues for the patient.Through follow-up we learned that for 2 lenses, parts were torn off after advancement trough the cartridge and they were removed and replaced.Two (2) further lenses were scratched or cracked by the cartridges, where the back haptic goes into the optic after pushing through the inserter.One of these lenses was removed while the second remains implanted without issues for the patient.No further information was provided.This report is for the second of two lenses with parts torn off the optic, which was removed and replaced from the patients eye.Separate reports will be submitted for the other devices involved.
|
| |
|
Additional Manufacturer Narrative
|
|
Section a2, a4 and a5: per regulation eu (b)(4) (general data protection regulation), patient identifiers were not collected or recorded and therefore are not available.Section d6a - implant date: not applicable.The cartridge is not an implantable device.Section d6b - explant date: not applicable.The cartridge is not an implantable device.Section e1 - last name: unknown, as information was requested but not provided section e1: email address: unknown/not provided, as information was asked but it was not provided.Section e1 - telephone number: (b)(6).Section h3 - the cartridge was not returned for evaluation.Therefore, a failure analysis of the complaint device could not be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review and possible product return and evaluation, if there is any further relevant information a supplemental medwatch will be filed.Attempts have been made to obtain the missing information.However, to date, no response has been received.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected information: jjsv has reviewed the provided information and determined that the reported issue is related to the emerald handpiece rather then the emerald cartridge.Therefor section d4 has been updated with the new suspect product details.The emerald cartridge was added as concomitant product under d10.In addition, the medical device problem code was updated to 1261 - material fragmentation.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
|
| |
|
Search Alerts/Recalls
|
|
|