• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARD® RENAL SHEATH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARD® RENAL SHEATH Back to Search Results
Model Number 992081
Medical Device Problem Code Positioning Failure (1158)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/06/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
 
Event or Problem Description
It was reported that a physician stated in an online survey that they were not able to successfully place the working sheath.The physician was not able to successfully dilate the nephrostomy tract for the device 992081 - x-force® 8mm x 17cm renal sheath, clear, sterile because the access was difficult.Per additional information received from ibc on 14jun2023, stated that the customer would like to update the initial reported event.The corrected event was that the physician stated in an online survey that they were not able to successfully place the working sheath.The physician was not able to successfully place the working sheath for the device 992081 - x-force® 8mm x 17cm renal sheath, clear, sterile because the access was difficult.
 
Event or Problem Description
It was reported that a physician stated in an online survey that they were not able to successfully place the working sheath.The physician was not able to successfully dilate the nephrostomy tract for the device 992081 - x-force® 8mm x 17cm renal sheath, clear, sterile because the access was difficult.As per additional information received from ibc on 14jun2023, stated that the customer would like to update the initial reported event.The corrected event was that the physician stated in an online survey that they were not able to successfully place the working sheath.The physician was not able to successfully place the working sheath for the device 992081 - x-force® 8mm x 17cm renal sheath, clear, sterile because the access was difficult.
 
Additional Manufacturer Narrative
The reported event is inconclusive as no sample was returned for evaluation.Although an exact root cause could not be determined a potential root cause could be beveled tip too blunt.The dhr review is not required as no lot number was reported.The instructions for use were found adequate and states the following: precautions: · only a physician who has an understanding of the clinical applications, technical principles and risks associated with balloon dilation of a nephrostomy tract should use this device.Directions for use: establishment of a percutaneous nephrostomy tract and renal dilation should always be performed under fluoroscopic control.1.Percutaneous access is established and a guidewire passed into a secure location within the intra-renal system.If feasible, the guidewire should be passed down the ureter and into the bladder.2.Dilation of the tract to the desired size should be performed using an x-force¿ nephrostomy balloon dilation catheter.3.Once the tract is established, the sheath is passed over the dilation device and positioned as desired.4.Remove the dilation instrument from within the sheath, maintaining both the sheath and guidewire in position.5.The guidewire may now be secured in the guidewire notch.6.The procedure may now be performed through the sheath.Section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® RENAL SHEATH
Common Device Name
RENAL SHEATH
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key17216195
Report Number1018233-2023-04760
Device Sequence Number16890397
Product Code LJE
UDI-Device Identifier00801741095023
UDI-Public(01)00801741095023
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K080944
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Study
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number992081
Device Catalogue Number992081
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/06/2023
Supplement Date Received by Manufacturer11/10/2023
Initial Report FDA Received Date06/28/2023
Supplement Report FDA Received Date11/14/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-