| Model Number |
663016 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
While no serious injury resulted in this event, if this malfunction recurred, it could cause or contribute to a serious injury or require medical or surgical intervention to preclude such.This event, therefore, is reportable per 21cfr part 803.The device is available for evaluation, though results are not available as of this report.Evaluation results will be submitted as they become available.
|
| |
|
Event or Problem Description
|
|
In this event it is reported that automatrix extra shaft assy broke during use in patient's mouth.Broken parts were retrieved, no injury occurred.
|
| |
|
Additional Manufacturer Narrative
|
|
On 8-2-2023: returned product; only the tip of the flexshaft assembly which has broken off from the rest of the assembly presumably due to coil deformation/weld failure.The date code of the flexshaft assembly could not be identified.No batch information provided in case therefore no dhr or retain evaluation can be conducted.
|
| |
|