|
The device was not received for analysis at the time of this report; therefore, no physical or visual analysis of the product could be performed.A review of the device history records of the specimen device does not present any indication of manufacturing defect or anomaly that could have impacted on the event as reported.During manufacturing and packaging of this product the operators are instructed to 100% visually inspect for any obvious defect prior to shipment.The history records indicate this product was final inspection tested at lake region medical and was determined to be acceptable.As described in the device instructions for use (dfu): ensure a catheter is properly placed in the ventricle or other intended treatment area.Carefully remove the safari2tm guidewire from the hoop by grasping the proximal end of the j-straightener separating the straightener from the hoop.After inspection of the safari2tm guidewire, advance the j-straightener over the distal section of the safari2tm guidewire to straighten the curve.Insert the safari2tm guidewire into the hub of the catheter with the aid of the j-straightener.Carefully advance the distal tip of the safari2tm guidewire into the lumen of the catheter.Remove the safari2tm guidewire j-straightener by withdrawing it over the length of the safari2tm guidewire.Advance the safari2tm guidewire through the catheter under fluoroscopic guidance to remove the guidewire from the treatment area: a catheter must be advanced into the treatment area over the safari2 guidewire prior to withdrawing the wire.The safari2 guidewire is pulled back completely into the catheter.The safari2 guidewire may then be withdrawn from the catheter as indicated in the device instructions for use warnings: when advancing or removing the guidewire, always use fluoroscopic guidance with radiographic equipment that provides high-resolution images.Never position the guidewire blindly, as this may result in misplacement, dissection, or perforation.Exercise care in handling of the guidewire during a procedure to reduce the possibility of accidental breakage, bending, kinking, or coil separation.Resulting guidewire fractures might require additional percutaneous intervention or surgery.Never advance the guidewire against the resistance without first determining the reason for resistance under fluoroscopy.Excessive force against resistance may result in damage to the catheter or vessel/organ.Care should be taken when advancing a guidewire after device deployment.At this time, it is not possible to assign a definitive root cause for the event as reported.The patient information was not provided.If there is any further relevant information received, a follow up medwatch report will be submitted.
|