• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR NC TREK CORONARY DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR NC TREK CORONARY DILATATION CATHETER Back to Search Results
Model Number 1012448-08
Medical Device Problem Code Device Markings/Labelling Problem (2911)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/08/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during opening of the package of a 2.75x8mm nc trek balloon dilatation catheter (bdc), the outer package labeling was noted to be inconsistent with the pouch labeling.Another same size nc trek balloon was used to complete the procedure.There was no device use or patient involvement.There was no clinically significant delay in the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.
 
Additional Manufacturer Narrative
A visual inspection was performed on the returned device.The reported complaint was confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation determined the reported complaint appears to be related to operational context.All currently available evidence points to the placement of the 2.75 x 15 mm nc trek bdc in the chipboard box for a 2.75 x 8 mm nc trek bdc having occurred at the account.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device as the respective lots were manufactured approximately ten months apart on different production lines and were never at the distribution centers at the same time, which indicates the two units involved in this complaint never came into contact with one another during production.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC TREK CORONARY DILATATION CATHETER
Common Device Name
CORONARY DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR COSTA RICA, REG # 3009564766
52 calle 3 b31 coyol free zone
el coyol alajuela
CS  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key17220692
Report Number2024168-2023-06908
Device Sequence Number6176803
Product Code LOX
UDI-Device Identifier08717648151804
UDI-Public08717648151804
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K103153
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2023
Device Model Number1012448-08
Device Catalogue Number1012448-08
Device Lot Number10104G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/08/2023
Supplement Date Received by Manufacturer08/17/2023
Initial Report FDA Received Date06/28/2023
Supplement Report FDA Received Date08/28/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/04/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-