| Model Number |
4068210 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
Concussion (2192)
|
| Date of Event |
06/08/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
It was identified that the screw which holds the retaining sleeve in place was missing.This allowed the retaining sleeve to move from its position and following the retaining segment to come out from the spring arm ¿ light bracket connection.As a result the light head detached and fell off.The missing screw was traced to an installation failure during previous service events of the light system.The light head was re-installed by the use of a new retaining segment and sleeve with screw.Based on this, no further actions are necessary.
|
| |
|
Event or Problem Description
|
|
The customer alleged that the room was being cleaned after a case.The cleaning technician adjusted light up to clean table, a few seconds later the light head fell hitting the cleaning tech in the head and then light head fell to floor.This report was filed in our complaint handling system as complaint #(b)(4).
|
| |
|