| Model Number |
TMT6B10 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Unspecified Infection (1930); Osteolysis (2377)
|
| Date of Event |
05/11/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Zimvie complaint number (b)(4).G4: pma/510(k) number: k132258, k113753 and k112160.
|
| |
|
Event or Problem Description
|
|
It was reported that the implant at tooth 15 was removed due to peri-implantitis.Patient would have to return to place a new implant.
|
| |
|
Additional Manufacturer Narrative
|
|
Zimmer biomet complaint number (b)(4).The following fields have been updated: b4: date of this report b5: describe event or problem d9: device availability g3: date received by manufacturer g6: type of report h1: type of reportable event h2: follow up type h3: device evaluated by manufacturer h4: device manufacturer date h6: adverse event problem h10: additional narrative dhr review was completed for the subject lot number 1230334.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Sterilization record was reviewed and verified to have passed all sterilization activities with no issues or nonconformities identified.Complaint history review was performed for the reported lot number 1230334 for similar events and no other complaint was identified.Review completed utilizing keywords: dental : medical : peri-implantitis a summary investigation has been completed for peri-implantitis events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for these events have not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of a manufacturing or design defect that might lead to peri-implantitis occurring and escaping the available detections has been assessed and found remote and almost nonexistent.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.H3 other text : summary investigation.
|
| |
|
Event or Problem Description
|
|
No further event information is available at the time of this report.
|
| |
|