Mechanical Problem (1384)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
In this event it is reported that a x-smart iq handpiece stopped running during use.The outcome of this event is unknown as of this mdr.Further information requested.
|
| |
|
Additional Manufacturer Narrative
|
|
There has been a previous report received where this malfunction has caused file separation.Since separation of a file could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information received that there was no injury in this event.Investigation results: received = 1x x-smart iq handpiece h105900000000 sn: (b)(6).X-smart iq handpiece sav other start / stop button intermittent defect.Replaced 1x iq handpiece h105900000000.
|
| |
|