• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL LLC AC3 OPTIMUS IABP NA/EMEA; SYSTEM, BALLOON, INTRA-AORTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL LLC AC3 OPTIMUS IABP NA/EMEA; SYSTEM, BALLOON, INTRA-AORTIC Back to Search Results
Model Number IPN917285
Medical Device Problem Code Communication or Transmission Problem (2896)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/02/2023
Type of Reportable Event Malfunction
Event or Problem Description
Reported as "systolic values and diastolic values reading as same number".The patient was a transfer from an outside facility with a clotted central lumen on a competitor iab.Alternate arterial access obtained and transduced.Troubleshooting included repositioning the patient, verifying the transducer signal and connections, rebooting iabp screen, power cycling the pump, and adjusting the timing.None of these measures resolved the issue.As a result, the pump was switched out for a different pump.The 2nd pump was manually timed with only minimal improvement.Iab volume reduction attempted with no improvement.No patient harm or injury.The patient status is reported as "fine".
 
Additional Manufacturer Narrative
(b)(4).
 
Additional Manufacturer Narrative
(b)(4).The reported complaint of "systolic values and diastolic values reading as same number" was confirmed based on the customer supplied pictures.The attached pictures show the diastolic and systolic pressures matching and improvement after troubleshooting.No iabp part was returned for investigation.Based on a review of the device history record (dhr), the product met specification upon release; however, the specifications were not met during the complaint investigation due to the incorrect hemodynamics.The root cause of the complaint was undetermined.No further action required at this time.The reported complaint will be monitored for any developing trends.Other remarks: n/a.Corrected data: n/a.
 
Event or Problem Description
Reported as "systolic values and diastolic values reading as same number".The patient was a transfer from an outside facility with a clotted central lumen on a competitor iab.Alternate arterial access obtained and transduced.Troubleshooting included repositioning the patient, verifying the transducer signal and connections, rebooting iabp screen, power cycling the pump, and adjusting the timing.None of these measures resolved the issue.As a result, the pump was switched out for a different pump.The 2nd pump was manually timed with only minimal improvement.Iab volume reduction attempted with no improvement.No patient harm or injury.The patient status is reported as "fine".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AC3 OPTIMUS IABP NA/EMEA
Common Device Name
SYSTEM, BALLOON, INTRA-AORTIC
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL INC.
16 elizabeth drive
chelmsford MA 01824
Manufacturer Contact
kevin don bosco
3015 carrington mill blvd
morrisville 27560
MDR Report Key17225513
Report Number3010532612-2023-00354
Device Sequence Number17908306
Product Code DSP
UDI-Device Identifier10801902172065
UDI-Public10801902172065
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K162820
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberIPN917285
Device Catalogue NumberIAP-0700
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/03/2023
Supplement Date Received by Manufacturer07/21/2023
Initial Report FDA Received Date06/29/2023
Supplement Report FDA Received Date07/24/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexFemale
-
-