|
The facility reported that during the procedure, it was noticed that the distal end of the forceps (ins-0372 always-on tip tracked forceps, 1.8mm od, serrated cup) appeared "very rough" on the endoscopic image.The customer was concerned that this would cause scope damage, especially when flexed.However, this had no effect on the ability to open, close or use the forceps.The physician was able to successfully complete the intended procedure after switching to a new instrument.There was no procedural delay and there was no injury to the patient or user as a result of the reported issue.
|