• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LEMAITRE VASCULAR, INC. TUFTEX OVER-THE-WIRE EMBOLECTOMY CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LEMAITRE VASCULAR, INC. TUFTEX OVER-THE-WIRE EMBOLECTOMY CATHETER Back to Search Results
Model Number 1651-44
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/29/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
We have not received the device for investigation.Hence, we could not conclusively determine the root cause of the reported incident.Due to the nature of the product, balloon ruptures are an expected risk when using this product.As stated in the ifu: as with all catheterization procedures, complications may occur.These may include but are not limited to: infection, local hematomas, intimal disruption, arterial dissection, perforation and rupture, hemorrhage, arterial thrombosis, distal emboli of blood clots or arteriosclerotic plaque, air embolus, aneurysms, arterial spasms, arteriovenous fistula formation, balloon rupture or tip separation with fragmentation and distal embolization.The possibility of balloon rupture must be taken into account when considering the risks involved in the catherization procedure.The lot history record for the device was reviewed, no issues were found during manufacturing or packaging that would cause or contribute to the reported event.Further, we have not received any other complaints of a similar nature for devices from this lot.
 
Event or Problem Description
It was reported the tuftex over-the-wire embolectomy catheter balloon ruptured while attempting to remove it from the vessel.No injury reported to patient.The operation was completed successfully using another tuftex over-the-wire embolectomy catheter.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TUFTEX OVER-THE-WIRE EMBOLECTOMY CATHETER
Common Device Name
EMBOLECTOMY CATHETER
Manufacturer (Section D)
LEMAITRE VASCULAR, INC.
63 second ave
burlington 01803
Manufacturer (Section G)
LEMAITRE VASCULAR, INC
63 second ave
burlington 01803
Manufacturer Contact
peter song
63 second ave
burlington 01803
7814251683
MDR Report Key17227160
Report Number1220948-2023-00101
Device Sequence Number7097611
Product Code DXE
UDI-Device Identifier00840663100675
UDI-Public(01)00840663100675
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K992368
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/28/2024
Device Model Number1651-44
Device Catalogue Number1651-44
Device Lot NumberXOT1125
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/30/2023
Initial Report FDA Received Date06/29/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/21/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-