Model Number NS7TCDL174HS
Medical Device Problem Code Patient Device Interaction Problem (4001)
Health Effect - Clinical Code Heart Block (4444)
Date of Event 06/01/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a patient underwent an atrioventricular nodal reentry tachycardia (avnrt) ablation procedure with a navistar¿ electrophysiology catheter.The patient suffered a heart block and required prolonged hospitalization.It was reported that a supra-hisian block occurred during avnrt procedure.Ablation point distant 17 mm from the his cloud.The patient stayed one day in the hospital waiting to recover the av node.This adverse event discovered during use of biosense webster products.The physician¿s opinion on the cause of this adverse event was that it was procedure related.The procedure was stopped, and the patient was kept under observation for one night in the hospital to check if the sinus rhythm was back to normal.It was reported that the outcome was ¿probably improved or unchanged¿.
 
Additional Manufacturer Narrative
If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster, inc., or its employees that the report constitutes an admission that the product, biosense webster, inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Manufacturer's reference number: (b)(4).
 
Additional Manufacturer Narrative
Additional information received indicates that the device is not available for return, therefore no product investigation can be performed, and the customer complaint cannot be confirmed.A manufacturing record evaluation was performed for the finished device 30730853m number, and no internal action related to the complaint was found during the review.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAVISTAR¿ ELECTROPHYSIOLOGY CATHETER
Common Device Name
CATHETER, ELECTRODE RECORDING
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER INC (JUAREZ)
circuito interior norte
1820parque industrial salvacar
juarez 32599
MX   32599
Manufacturer Contact
kate karberg
31 technology dr
irvine, CA 92618
3035526892
MDR Report Key17228969
Report Number2029046-2023-01426
Device Sequence Number7207371
Product Code DRF
UDI-Device Identifier10846835000597
UDI-Public10846835000597
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
P990025
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/01/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/13/2025
Device Model NumberNS7TCDL174HS
Device Catalogue NumberNS7TCDL174HS
Device Lot Number30730853M
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/01/2023
Supplement Date Received by Manufacturer08/08/2023
Initial Report FDA Received Date06/29/2023
Supplement Report FDA Received Date09/01/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/14/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SMARTABLATESUPERKIT (G+P+R)-WW
Outcome Attributed to Adverse Event Hospitalization; Life Threatening;
Patient SexUnknown
-
-