Model Number 2975-29-075
Medical Device Problem Codes Naturally Worn (2988); Scratched Material (3020)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/02/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the reclaim neck trials were not staying in position when surgeon was doing final trailing.There was no delay in procedure and all damaged trials were recovered.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: the device associated with this report was returned to depuy synthes for evaluation.Visual examination found the device worn and heavily scratched on the metal grooves on the shaft.The reported allegation can be confirmed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RECLAIM PROX TRIAL SHAFT 75
Common Device Name
RECLAIM INSTRUMENTS : FEMORAL TRIALS
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY IRELAND 9616671
loughbeg, ringaskiddy co.
cork
EI  
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key17231803
Report Number1818910-2023-13443
Device Sequence Number7098169
Product Code HWT
UDI-Device Identifier10603295156260
UDI-Public10603295156260
Combination Product (Y/N)N
Initial Reporter StateAK
Initial Reporter CountryUS
PMA/510(K) Number
K903084
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number2975-29-075
Device Catalogue Number297529075
Device Lot NumberSO2025064
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/28/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/21/2023
Supplement Date Received by Manufacturer07/05/2023
Initial Report FDA Received Date06/29/2023
Supplement Report FDA Received Date07/05/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/03/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-