MAUDE Adverse Event Report: INTUITIVE SURGICAL, INC SUREFORM; SUREFORM 45 RELOAD BLUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTUITIVE SURGICAL, INC SUREFORM; SUREFORM 45 RELOAD BLUE Back to Search Results
Model Number 48345B-01
Medical Device Problem Code Failure to Fire (2610)
Health Effect - Clinical Code Hemorrhage/Blood Loss/Bleeding (1888)
Date of Event 05/31/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during a da vinci-assisted pulmonary segmentectomy procedure, while stapling across the bronchus with a sureform 45 curved-tip stapler instrument loaded with a green sureform 45 curved-tip stapler reload, there was a partial fire, and the end of the staple line was not sealed.The staples from the end of the reload did not deploy.There was an unspecified amount of bleeding due to the gap in the staple line.The surgeon switched to a blue sureform 45 curved-tip stapler reload, and the same issue occurred.The surgeon switched to a laparoscopic stapler and continued the staple line, with no further issues.The stapler had fired an unspecified number of loads successfully, prior to this issue.The surgeon used an intraoperative bronchoscope to check the staple line.Both the green and blue reloads had exposed blades.Per the reporter, the surgeon downsized to a blue reload after using the green reload, due to the thickness of the target tissue at the end of the intended staple line.The target tissue was reportedly in a normal state, and the patient was reportedly in a normal condition following the procedure.Information regarding patient demographics, relevant testing, and medical history was requested, however the reporter was not able to provide that information.
 
Additional Manufacturer Narrative
Based on the customer's claim against the product, an investigation is in progress to determine the cause of this reported event.If additional information is obtained, a follow-up mdr will be submitted.Per a review of the site's system logs for the reported procedure date, no related system errors occurred during the surgical procedure that would have likely caused or contributed to the reported complaint.A review of the advanced instrument log for the instrument associated with this event was performed by an intuitive surgical, inc.(isi) failure analysis engineer (fae).Per the fae, the logs show the instrument was installed on the system 3 times, and it fired 3 reloads (1 white, 1 green, 1 blue, in that order).On install 1, the first clamp was successful, and the firing was completed, with no pauses for compression.On install 2, the first clamp was successful, but it was followed by a firing failure.The firing stopped at ~89% completion, after a duration of 57 seconds.The maximum firing torque was recorded at 699 n.On install 3, the first clamp was successful, but it was followed by a firing failure.This firing stopped at ~78% completion, after a duration of 26 seconds.The maximum firing torque was recorded at 694 n.The instrument was then removed and was not used again in the procedure.There were no stapler related errors in the logs.
 
Additional Manufacturer Narrative
Intuitive surgical, inc.(isi) received the blue sureform 45 curved-tip stapler reload associated with this complaint, and a failure analysis (fa) investigation was completed.Fa replicated and confirmed the customer reported complaint of a partial fire, with an exposed blade as the primary failure finding.The reload was found to have the knife exposed within the knife track.It was also found to have a firing failure, based on a log review during the in-house inspection.Commonly, the probable root cause of exposed components is typically attributed to misuse, such as firing across a staple line or other obstructions.Additionally, the reload was found to have mechanical indentation damage to the blade and cartridge damage, with a missing piece approximately 0.078" x 0.03" in size, that was not returned with the reload.This damage is commonly attributed to mishandling/misuse.Isi also received the sureform 45 curved-tip stapler instrument associated with this complaint, and a failure analysis (fa) investigation was completed.Fa confirmed, but did not replicate, the customer reported complaint of a partial fire, with a firing failure as the primary finding.The instrument was found to have firing failures, based on a review of the device logs, with an indeterminable probable root cause.The instrument was placed and driven on an in-house system, where it passed initialization and moved intuitively, with full range of motion in all directions.The jaws opened and closed properly, and the instrument clamped, fired, and unclamped, without any issues.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUREFORM
Common Device Name
SUREFORM 45 RELOAD BLUE
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer (Section G)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer Contact
izabel nielson
3410 central expressway
santa clara, CA 
4085232100
MDR Report Key17236361
Report Number2955842-2023-16717
Device Sequence Number5754891
Product Code GDW
UDI-Device Identifier10886874117085
UDI-Public(01)10886874117085(10)NA
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
K183224
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 05/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number48345B-01
Device Catalogue Number48345B
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/29/2023
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/31/2023
Supplement Date Received by Manufacturer06/30/2023
Initial Report FDA Received Date06/30/2023
Supplement Report FDA Received Date07/27/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/02/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DA VINCI INSTRUMENTS AND ACCESSORIES
Patient SexUnknown
-
-