C.R. BARD, INC. (BASD) -3006260740 M.R.I. LOW-PROFILE IMPLANTABLE PORT, HICKMAN SINGLE-LUMEN, 6.6F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR Back to Search Results
Model Number 0603880
Medical Device Problem Codes Disconnection (1171); Difficult to Flush (1251); Blocked Connection (2888); Migration (4003)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 06/06/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that four months and nineteen days post port placement procedure, the doctor allegedly felt a difficulty in injecting the product while inserting the needle for the injection of an immunotherapy treatment.It was further reported that the implantable chamber catheter allegedly disconnected from the chamber box.Reportedly, the catheter was found to be migrated between the superior vena cava and the right ventricle.The current status of the patient is unknown.
 
Additional Manufacturer Narrative
H10: as the lot number for the device was provided, a review of the device history records will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.H10: d4 (expiry date: 05/2025).H11: section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Additional Manufacturer Narrative
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.Therefore, the investigation is inconclusive for the reported failure as no objective evidence was provided for review.The definitive root cause could not be determined based upon available information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H10: d4 (expiration date: 05/2025).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that four months and nineteen days post port placement, the doctor allegedly felt a difficulty in injecting the product while inserting the needle for the injection of an immunotherapy treatment.It was further reported that the implantable chamber catheter allegedly disconnected from the chamber box.Reportedly, the catheter was found to be migrated between the superior vena cava and the right ventricle.The current status of the patient is unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
M.R.I. LOW-PROFILE IMPLANTABLE PORT, HICKMAN SINGLE-LUMEN, 6.6F
Common Device Name
PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key17236506
Report Number3006260740-2023-02711
Device Sequence Number5806952
Product Code LJT
UDI-Device Identifier00801741025846
UDI-Public(01)00801741025846
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K924250
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date (Section B) 09/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0603880
Device Catalogue Number0603880
Device Lot NumberREGQ2937
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/07/2023
Supplement Date Received by Manufacturer11/14/2023
Initial Report FDA Received Date06/30/2023
Supplement Report FDA Received Date11/14/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/28/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age55 YR
Patient SexFemale
Patient Weight47 KG
-
-