• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC TRACKER 9733533XOM ENT INSTRUMENT; NEUROLOGICAL STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC TRACKER 9733533XOM ENT INSTRUMENT; NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Model Number 9733533XOM
Medical Device Problem Codes Imprecision (1307); Device Sensing Problem (2917)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/07/2023
Type of Reportable Event Malfunction
Event or Problem Description
Medtronic received information regarding a navigation system used intraoperatively in a functional endoscopic sinus surgery (fess) in which there was patient involvement.It was reported that a patient was registered without issue and navigation looked accurate in verify registration task.The straight suction was verified without issue but the system was not able to recognize that their curved suction was attempting to verify on the same instrument tracker.The instrument tracker was swapped and were then able to verify both the straight and curved suctions without issue.After resolving that issue, the surgeon reportedly felt inaccurate.Navigational accuracy was not able to be re-established.Navigation was aborted.Surgical delay was reported as less than one hour and there was no impact to patient outcome.
 
Additional Manufacturer Narrative
H3) the tracker was returned to the manufacturer for analysis.After functional testing and visual/physical examination, the reported issue was not confirmed.The returned tracker was found to be fully functional when connected to a known good system.The tracker displayed normal tracking and divot error.Codes: b01, c19, d14.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRACKER 9733533XOM ENT INSTRUMENT
Common Device Name
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key17237367
Report Number1723170-2023-01082
Device Sequence Number12528682
Product Code HAW
UDI-Device Identifier00763000588342
UDI-Public00763000588342
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K001284
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/22/2024
Device Model Number9733533XOM
Device Catalogue Number9733533XOM
Device Lot Number220623A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/15/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/12/2023
Initial Report FDA Received Date06/30/2023
Date Device Manufactured06/23/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age61 YR
Patient SexMale
Patient Weight79 KG
-
-