| Model Number |
15 |
| Medical Device Problem Code |
Failure to Deliver Shock/Stimulation (1133)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/03/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer contacted stryker to report that their device failed to provide defibrillation therapy during intervention on a patient.This issue is patient related; however there was no adverse patient outcome reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Stryker contacted the customer to request additional information on the patient.No response has been received from the customer.Patient fields in which information is not provided were intentionally left blank.Stryker service representative performed an initial evaluation of the customer¿s device and was not able to verify the reported issue.Proper device operation was observed through functional and performance testing.The device was subsequently returned to the customer.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer provided stryker with the available patient information.Patient fields in which information is not provided were intentionally left blank.
|
| |
|
Event or Problem Description
|
|
The customer contacted stryker to report that their device failed to provide defibrillation therapy during intervention on a patient.This issue is patient related; however there was no adverse patient outcome reported.
|
| |
|