| Model Number |
MOSAIQ |
| Medical Device Problem Code |
Computer Software Problem (1112)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/26/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported kv portal images were taken but did not show in the treatment chart.
|
| |
|
Additional Manufacturer Narrative
|
|
The manufacturer's investigation is on-going and further information will be provided once the investigation has completed.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation was completed by conducting a thorough evaluation of the product and the reported information.The logs show that after the kv images were taken, the ct was returned due to an acquisition mode discrepancy where the preset was not auto-selected in xvi, therefore mosaiq did not record the field in the treatment chart.Mosaiq did not have a malfunction and was working as designed and intended.This issue is limited to recording the images in the treatment chart and did not affect clinical patient setup.
|
| |
|