| Model Number |
2700TFX |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Codes |
Heart Failure/Congestive Heart Failure (4446); Insufficient Information (4580)
|
| Date of Event |
06/07/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Additional manufacturer narrative: surgical/percutaneous intervention is indicated or performed, or harm occurred due to the device, or there is a device malfunction that could cause or contribute to a serious injury.This event is considered a serious injury.The device was not returned for evaluation, as it remains implanted.The investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient with a 29mm 2700tfx aortic valve underwent a valve-in-valve procedure after an implant duration of 6 years, 11 months due to severe aortic insufficiency.The tavr was performed with a 29mm 9755rsl transcatheter valve.The procedure was successful.The implanting team stated the valve performed as expected.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient with a 29mm 2700tfx aortic valve underwent a valve-in-valve procedure after an implant duration of 6 years, 11 months due to severe aortic insufficiency.The patient presented with heart failure.The tavr was performed with a 29mm 9755rsl transcatheter valve.The procedure was successful.The patient was stable post procedure and discharged on pod #1.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: additional manufacturer narrative: regurgitation which develops progressively over time can be due to a number of issues including patient related factors or structural valve deterioration.Structural valve deterioration (svd) can, and typically does, lead to chronic central leaks over a period of time.Svd is the most common reason for bioprosthesis explant and encompasses multiple failure modes, including calcification, noncalcific degeneration, dehiscence, cusp thickening or fibrosis, or a combination of these.Such failure modes may occur singularly or concomitantly.Tissue degeneration-related structural deterioration, either calcific or non-calcific, are common chronic failure modes for this type of bioprosthetic heart valve.Operational mechanical stress and biological factors are generally believed to be the major contributors to the non-calcific bioprosthetic tissue degeneration.A definitive root cause cannot be conclusively determined; however, patient factors likely caused or contributed.The device history record (dhr) review was completed and this device passed all manufacturing and sterilization inspections prior to release for distribution.There were no issues identified that would have impacted this event.
|
| |
|