MAUDE Adverse Event Report: LAKE REGION MEDICAL ZIPWIRE HYDROPHILIC GUIDEWIRE; UROLOGICAL CATHETER AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LAKE REGION MEDICAL ZIPWIRE HYDROPHILIC GUIDEWIRE; UROLOGICAL CATHETER AND ACCESSORIES Back to Search Results
Model Number M0066802231
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 05/26/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This mdr report is being filed as a malfunction to align with the documented event recorded on mdr report # mw5118126.Note, the initial reporter elected to remain confidential; therefore, many of the fields within this report have been left blank.The device was not received for evaluation at the time of this report; therefore, no physical analysis of the device can be performed.A review of the manufacturing processes indicates the production operators are instructed to 100% visually and tactilely inspect for any obvious defect, which includes a tactile examination of the entire length of each wire.In addition, during packaging of this product the operators are instructed to 100% visually inspect for any obvious defect prior to shipment.The history records indicate this product was final inspection tested at lake region medical and was determined to be acceptable.As noted in the device instructions for use (dfu) warnings, do not manipulate, advance and/or withdraw the zipwire hydrophilic guidewire through a metal cannula or needle.Manipulation, advancement, and/or withdrawal through a metal device may result in destruction and/or separation of the outer polymer jacket requiring retrieval.If a needle is used for initial placement, a plastic entry needle is recommended when using the zipwire hydrophilic guidewire.Extreme caution should be observed when used with a one-wall puncture needle.The dfu precautions also indicate, the zipwire hydrophilic guidewire should be advanced through the scope using short, deliberate 2-3cm movements to prevent inadvertent damage to the device or patient.Due to variations in certain catheter tip diameters, abrasion of the hydrophilic coating may occur during manipulation.If any resistance is felt during introduction of the catheter, it is advisable to stop using such catheters.At this time, it is not possible to assign a definitive root cause for the event as reported.If there is any further relevant information provided, a follow up medwatch report will be submitted.
 
Event or Problem Description
Per medwatch, it was reported that: safety wire was removed.Inspection of glidewire at the end of the case showed 2 millimeter(s) (mm) tip of glidewire broke off.Note from distributor received (b)(6), 2023: we will not be able to send a follow up on this complaint since initial reporter and health care facility are unknown.The device is returning; however, it has not been received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZIPWIRE HYDROPHILIC GUIDEWIRE
Common Device Name
UROLOGICAL CATHETER AND ACCESSORIES
Manufacturer (Section D)
LAKE REGION MEDICAL
340 lake hazeltine drive
chaska MN 55318
Manufacturer (Section G)
VENUSA DE MEXICO S. DE R.L. DE C.V.
calle hertz 1525
parque industrial a j bermudez
ciudad juarez, chihuahua 32470
MX   32470
Manufacturer Contact
victoria foss
340 lake hazeltine drive
chaska, MN 55318
9526418506
MDR Report Key17239266
Report Number9680001-2023-00108
Device Sequence Number7207137
Product Code EYA
UDI-Device Identifier10816349011839
UDI-Public10816349011839
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 06/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2025
Device Model NumberM0066802231
Device Lot NumberJRZ7140466
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/14/2023
Initial Report FDA Received Date06/30/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/13/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-