• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL LLC FIBEROPTIX ULTRA 8 IAB: 8FR 40CC; SYSTEM, BALLOON, INTRA-AORTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL LLC FIBEROPTIX ULTRA 8 IAB: 8FR 40CC; SYSTEM, BALLOON, INTRA-AORTIC Back to Search Results
Model Number IPN039787
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code Needle Stick/Puncture (2462)
Date of Event 06/06/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Qn#(b)(4).
 
Event or Problem Description
It was reported that while in use the iabp had "frequent possible helium loss 3 alarms".The iab was noted to have a severe kink it in.As a result, a 2nd iab was inserted in the opposite femoral artery.No patient harm or injury.The patient status is reported as "fine".
 
Event or Problem Description
It was reported that while in use the iabp had "frequent possible helium loss 3 alarms".The iab was noted to have a severe kink it in.As a result, a 2nd iab was inserted in the opposite femoral artery.No patient harm or injury.The patient status is reported as "fine".
 
Additional Manufacturer Narrative
(b)(4).The reported complaint for iab "possible helium loss 3 alarms" was confirmed based on the customer photos provided with the complaint report.The product was not returned for investigation.A device history record (dhr) review was conducted for the lot number with no relevant findings.The device passed all manufacturing specifications prior to release.The occurrence rate of the reported complaint was within acceptable risk limits.The root cause of the complaint was undetermined.No further action required at this time.The reported complaint will be monitored for any developing trends.A full investigation of the sample will be completed, once the product is returned.Other remarks: n/a.Corrected data: n/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FIBEROPTIX ULTRA 8 IAB: 8FR 40CC
Common Device Name
SYSTEM, BALLOON, INTRA-AORTIC
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL INC.
16 elizabeth drive
chelmsford MA 01824
Manufacturer Contact
kevin don bosco
3015 carrington mill blvd
morrisville 27560
MDR Report Key17240594
Report Number3010532612-2023-00365
Device Sequence Number14767406
Product Code DSP
UDI-Device Identifier10801902145717
UDI-Public10801902145717
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K021462
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2024
Device Model NumberIPN039787
Device Catalogue NumberIAB-05840-LWS
Device Lot Number18F23A0026
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/06/2023
Supplement Date Received by Manufacturer07/26/2023
Initial Report FDA Received Date06/30/2023
Supplement Report FDA Received Date07/28/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/13/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexMale
Patient Weight31 KG
-
-