| Model Number |
IPN039787 |
| Medical Device Problem Code |
Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
Needle Stick/Puncture (2462)
|
| Date of Event |
06/06/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Qn#(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that while in use the iabp had "frequent possible helium loss 3 alarms".The iab was noted to have a severe kink it in.As a result, a 2nd iab was inserted in the opposite femoral artery.No patient harm or injury.The patient status is reported as "fine".
|
| |
|
Event or Problem Description
|
|
It was reported that while in use the iabp had "frequent possible helium loss 3 alarms".The iab was noted to have a severe kink it in.As a result, a 2nd iab was inserted in the opposite femoral artery.No patient harm or injury.The patient status is reported as "fine".
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).The reported complaint for iab "possible helium loss 3 alarms" was confirmed based on the customer photos provided with the complaint report.The product was not returned for investigation.A device history record (dhr) review was conducted for the lot number with no relevant findings.The device passed all manufacturing specifications prior to release.The occurrence rate of the reported complaint was within acceptable risk limits.The root cause of the complaint was undetermined.No further action required at this time.The reported complaint will be monitored for any developing trends.A full investigation of the sample will be completed, once the product is returned.Other remarks: n/a.Corrected data: n/a.
|
| |
|