| Model Number |
SYS-4000 |
| Medical Device Problem Code |
Display or Visual Feedback Problem (1184)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/15/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The playback data from the procedure was analyzed.It was concluded the instrument tip and vpads locations reported by the localization system were moving around far more than is expected during normal use.The software was displaying the location of the instrument, as reported.The system cart was returned to the manufacturer for investigation.In a controlled setting, the system continued to report abnormal levels of movement with several instruments.After replacing the system interface unit and field generator, the system behaved as expected.The device was repaired to specification and returned to the customer.Veran will continue to monitor field performance for this device.This event is being reported as a corrective action following a retrospective review of complaints in response to an observation from an external inspection.
|
| |
|
Event or Problem Description
|
|
The physician reported that the screen was shaking when navigating.The computed tomography (ct) and virtual views had the cross-hairs jumping around even when the instrument was held steady and not moving.The navigated procedure was aborted.This report is being submitted for the aborted navigation procedure while the patient was under anesthesia.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to correct h4: device manufacture date from the initial medwatch report.The correct device manufacture date is 24-nov-2020.
|
| |
|