• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: KARL STORZ SE & CO. KG LIGHT CABLE ASSEMBLY, 4.8 MM X 3000MM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

KARL STORZ SE & CO. KG LIGHT CABLE ASSEMBLY, 4.8 MM X 3000MM Back to Search Results
Model Number 495NE
Medical Device Problem Code Human-Device Interface Problem (2949)
Health Effect - Clinical Code Burn(s) (1757)
Date of Event 06/20/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The previous reportability decision for this event was reversed based on a retrospective review.We were notified a patient was burned using a 5mm scope and the 495ne light cable.We were told the surgeon demanded the largest size light cable even though it was recommended to use the smaller (495nd) light cable.We have requested further info from the customer regarding product, surgeon and more info regarding the incident, but have not been give additional details.The scope and light cable were not tagged for replacement so cannot be sent in.There was patient injury.In conclusion, the complaint is considered as reportable.Device was not returned to manufacturer.Internal karl storz complaint id: (b)(4).
 
Event or Problem Description
We were notified a patient was burned using a 5mm scope and the 495ne light cable.We were told the surgeon demanded the largest size light cable even though it was recommended to use the smaller (495nd) light cable.We have requested further info from the customer regarding product, surgeon and more info regarding the incident, but have not been give any additional details.The scope and light cable were not tagged for replacement so cannot be sent in.There was patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGHT CABLE ASSEMBLY, 4.8 MM X 3000MM
Common Device Name
LIGHT CABLE ASSEMBLY, 4.8 MM X 3000MM
Manufacturer (Section D)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
tuttlingen, 78532
GM  78532
MDR Report Key17242744
Report Number2020550-2023-00145
Device Sequence Number20510282
Product Code FCW
UDI-Device Identifier04048551168905
UDI-Public4048551168905
Combination Product (Y/N)N
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Importer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 06/30/2023
Report Date (Section F) 06/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number495NE
Device Catalogue Number495NE
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Distributor Facility Aware Date06/26/2023
Event Location Hospital
Date Report to Manufacturer06/30/2023
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date06/30/2023
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-