• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MO016 MO-MOBERLY NOVAPLUS INFANT HEEL WARMER SQUEEZE; INFANT HEEL WARMER (CHEMICAL HEAT PACK)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MO016 MO-MOBERLY NOVAPLUS INFANT HEEL WARMER SQUEEZE; INFANT HEEL WARMER (CHEMICAL HEAT PACK) Back to Search Results
Model Number V11460-010
Medical Device Problem Code Burst Container or Vessel (1074)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/01/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This complaint was forwarded to the manufacturing facility where it is currently still under investigation.A follow up report will be filed once the results have been completed.
 
Event or Problem Description
It has been reported that when a nurse attempted to activate a hot pack using normal pressure the pack came apart.The broken side was facing upwards and as it broke it burst forth its contents spraying onto the nurse and across two occupied bassinets.Some of the contents splashed onto the nurses hands and on her chest.The contents were immediately cleaned up without any interaction.Nurse is unaffected by the event.
 
Additional Manufacturer Narrative
Cardinal health has received an increase in burst/leaking complaints regarding the novalplus infant heel warmer with tape.Customer complaints have noted rupturing and bursting of the device when activation is attempted, making the device unusable and in many cases causing the contents inside the pouch to leak and contact clinicians, patients, and other individuals or surfaces in proximity.Samples returned to cardinal health from customers have been evidenced to have an incomplete top seal, resulting in the malfunction.There have been no reportable adverse events/injuries associated with v11460-010, lot v2s056.Cardinal health has halted distribution of the novalplus infant heel warmer w/tape for lot number v2s056 and initiated a voluntary recall on june 16, 2023.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NOVAPLUS INFANT HEEL WARMER SQUEEZE
Common Device Name
INFANT HEEL WARMER (CHEMICAL HEAT PACK)
Manufacturer (Section D)
MO016 MO-MOBERLY
808 w highway 24
moberly MO 65270
Manufacturer (Section G)
MO016 MO-MOBERLY
808 w highway 24
moberly MO 65270
Manufacturer Contact
patricia tucker
3651 birchwood dr.
waukegan, IL 60085
8478874151
MDR Report Key17243373
Report Number1423537-2023-00957
Device Sequence Number4795989
Product Code MPO
UDI-Device Identifier10192253061468
UDI-Public10192253061468
Combination Product (Y/N)N
Initial Reporter StateNH
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 07/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberV11460-010
Device Catalogue NumberV11460-010
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/06/2023
Supplement Date Received by Manufacturer06/06/2023
Initial Report FDA Received Date06/30/2023
Supplement Report FDA Received Date07/10/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age25 YR
Patient SexFemale
Patient Weight75 KG
-
-