Model Number PM3140
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Infection (1930); Pocket Erosion (2013)
Date of Event 06/13/2023
Type of Reportable Event Serious Injury
Event or Problem Description
Related manufacturer reference number: 2017865-2023-24888 related manufacturer reference number: 2017865-2023-24889 related manufacturer reference number: 2017865-2023-24890 it was reported that the patient presented in clinic for follow up.Upon review it discovered that the pacemaker and leads had eroded.The physician confirmed that there was an infection.The entire system was explanted, and a wound debridement surgery was performed.The patient was in stable condition post procedure.
 
Additional Manufacturer Narrative
Visual examination found no malfunctions.A device history record (dhr) review was performed, and review of the sterilization records confirmed normal sterilization cycles for the products.The device met specifications prior to leaving abbott manufacturing facilities.The product was returned, and visual inspection was normal.The cause of infection could not be traced to the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ALLURE CRT-P
Common Device Name
NO MATCH
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key17243465
Report Number2017865-2023-24887
Device Sequence Number17867073
Product Code NIK
UDI-Device Identifier05414734507899
UDI-Public05414734507899
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P030035
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/31/2022
Device Model NumberPM3140
Device Catalogue NumberPM3140
Device Lot NumberP000124622
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/26/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/13/2023
Supplement Date Received by Manufacturer07/10/2023
Initial Report FDA Received Date06/30/2023
Supplement Report FDA Received Date07/27/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/09/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-