| Model Number |
050-50758 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/23/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A peritoneal dialysis (pd) patient contacted fresenius technical support to report the liberty select cycler alarmed with an m31 air detected in cassette warning during drain 1 of 5.A fluid leak was subsequently discovered in step 9 of treatment complete.The patient was not connected during the incident.Fluid was found in both the pump module area and on the external portion of the cassette.It is unknown at which point in therapy the leak may have begun.The cause of the leak is unknown.Upon followup the clinic manager (cm) at the patient¿s clinic confirmed the reported event.The cm stated that patient did not develop any symptoms, adverse events, injuries, or require medical intervention as a result of the reported event.The cm could not confirm if the patient completed treatment on the day of the reported event.The cm confirmed the patient has continued using the cycler for their pd treatment without issue since.The cm confirmed that the cycler set was discarded and not available to be returned to the manufacturer for physical evaluation.No defect or damage to the cycler set cassette was noted upon removing it from the cycler.
|
| |
|
Additional Manufacturer Narrative
|
|
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: d1, d4, g4 pma/510(k)#, h4 plant investigation: the sample was not returned to the manufacturer and the lot number was not provided.A manufacturing review was performed on the products shipped to the patient for the three (3) month time frame which immediately preceded the event occurrence date.This review included the lot numbers for all fresenius liberty cycler sets shipped to this account within the selected time frame.The entire set of lots have been sold and distributed.There were no non-conformances or abnormalities identified during the manufacturing process which could be associated with the reported event.An investigation of the device history records (dhr) was conducted and confirmed that the results of the in-progress and final quality control (qc) testing met all requirements.The product lots involved met all specifications for release.A review of the dhr did not reveal a probable cause for the customer complaint.As a physical evaluation could not be performed, a definitive conclusion regarding the reported incident could not be reached and a cause could not be confirmed.
|
| |
|
Event or Problem Description
|
|
A peritoneal dialysis (pd) patient contacted fresenius technical support to report the liberty select cycler alarmed with an m31 air detected in cassette warning during drain 1 of 5.A fluid leak was subsequently discovered in step 9 of treatment complete.The patient was not connected during the incident.Fluid was found in both the pump module area and on the external portion of the cassette.It is unknown at which point in therapy the leak may have begun.The cause of the leak is unknown.Upon followup the clinic manager (cm) at the patient¿s clinic confirmed the reported event.The cm stated that patient did not develop any symptoms, adverse events, injuries, or require medical intervention as a result of the reported event.The cm could not confirm if the patient completed treatment on the day of the reported event.The cm confirmed the patient has continued using the cycler for their pd treatment without issue since.The cm confirmed that the cycler set was discarded and not available to be returned to the manufacturer for physical evaluation.No defect or damage to the cycler set cassette was noted upon removing it from the cycler.
|
| |
|