| Model Number |
0500416E |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Hemorrhage/Blood Loss/Bleeding (1888)
|
| Date of Event |
06/15/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A user facility registered nurse (rn) reported an internal dialyzer blood leak that occurred immediately in a patient¿s hemodialysis (hd) treatment.The blood leak was visually observed in the dialysate compartment of the dialyzer.The machine, a fresenius 2008k2 machine, alarmed appropriately with a blood leak alarm.A blood leak test strip was used and tested positive.No visible damage was observed on the dialyzer.After the machine alarmed, the treatment was paused.The patient¿s blood was not returned; estimated blood loss (ebl) was 50 ml.The rn confirmed there was no patient injury, no adverse effects were experienced, and no medical intervention was required as a result of the reported event.The patient completed treatment after being resetup with new supplies on a different machine.The complaint device was reportedly not available to be returned for manufacturer evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity.
|
| |
|
Additional Manufacturer Narrative
|
|
Plant investigation: the reported complaint was not confirmed as the complaint device was not returned for manufacturer evaluation.As no lot number was provided by the complainant, a delivery search was performed to identify all lot numbers with the reported catalog number shipped to the complainant in the three months prior to the occurrence date.There were no lot identified within this time frame.The search was then extended to find the last lot delivered with the reported catalog number.Again, there were no results; therefore, a lot history review or production record review could not be performed.A definitive conclusion regarding the complaint incident cannot be reached without physical examination of the actual device.Therefore, the complaint is not confirmed.Continuous improvement is of the utmost importance to fresenius medical care as we strive to provide dialysis products of the highest quality to our patients.Reports of leaking product are investigated both individually as complaints as well as via the non-conformance/capa program, in order to assess and improve our products and processes.Capas for vision systems and blood leak reduction are recent examples of leak related investigations directed at an overall reduction in dialyzer leaks.
|
| |
|
Event or Problem Description
|
|
A user facility registered nurse (rn) reported an internal dialyzer blood leak that occurred immediately in a patient¿s hemodialysis (hd) treatment.The blood leak was visually observed in the dialysate compartment of the dialyzer.The machine, a fresenius 2008k2 machine, alarmed appropriately with a blood leak alarm.A blood leak test strip was used and tested positive.No visible damage was observed on the dialyzer.After the machine alarmed, the treatment was paused.The patient¿s blood was not returned; estimated blood loss (ebl) was 50 ml.The rn confirmed there was no patient injury, no adverse effects were experienced, and no medical intervention was required as a result of the reported event.The patient completed treatment after being resetup with new supplies on a different machine.The complaint device was reportedly not available to be returned for manufacturer evaluation.
|
| |
|
Search Alerts/Recalls
|