Conclusion: medtronic cannot confirm or deny the complaint of leak from the luer cap as no product has been returned to date.A root cause of this occurrence cannot be determined without returned product.Review of the device history record found no abnormalities during manufacturing that would cause or contribute to the reported event.Complaints received from november 2020 through january 2022 for similar model numbers were reviewed and showed no trends warranting escalation related to this occurrence.If the product is returned, this pe will be reopened and the analysis and investigation will be updated.There were no adverse patient effects as a result of this occurrence.Medtronic has made its supplier aware of this occurrence and will continue to monitor for future occurrences and trends per the product quality meetings procedure.This regulatory report is being submitted as part of a retrospective review and remediation per (b)(4) as part of a capa action.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
Medtronic received information that during use of an eopa arterial cannula, the customer reported that they could not get the vented cap on the cannula to tighten and it leaked blood during the entire case.They estimate that patient lost 100ml of blood because of this leak, but cannula was not changed out and the case was completed successfully without injury to the patient other than blood loss. additional information: the cannula was not heated or cooled prior to use.It was asked but unknown whether a blood transfusion was required.
|