|
This device is classified as import for export, therefore 510k is not applicable.Model eg-1690k is available in the usa with a 510k number k131902.The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the operation channel (primary) stuck accessory/object.Based on the result, we concluded that it was caused due to the inadequate/insufficient reprocessing at the facility on the operation channel (primary).In addition, our technician confirmed that the insertion flexible tube buckled, the bending rubber cut, and the lcb (light carrying bundle) defective; however, these defects are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report "hr-rpt-0588(channel)" and/or the risk analysis results, it was evaluated to submit mdr.
|