Model Number 78220
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/06/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device evaluation summary: visual inspection of the unopened device shows evidence of damage / deformity.Reason for return was confirmed.Conclusion: after investigation at medtronic and viant, complaint is confirmed for damage to the cannula body.This complaint is the result of a manufacturing issue the cannula was caught in the seal of the pouch during ffs packaging.Review of the device history record found no abnormalities during manufacturing that would cause or contribute to the reported event.Complaints received from september 2020 through november 2021 for similar model numbers were reviewed and showed no trends warranting escalation related to this occurrence.There were no adverse patient effects as a result of this incidence.Medtronic has made its supplier aware and will continue to monitor for future occurrences and trends per the product quality meetings procedure.This regulatory report is being submitted as part of a retrospective review and remediation per d00953163 as part of a capa action.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event or Problem Description
Medtronic received information that prior to use of this cannula the customer noted that the cannula had a significant crimp.The issue was noted before the cannula was handed up to the sterile field.There was no patient involvement so no adverse effects occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EOPA 3D VENTED ARTERIAL CANNULA
Common Device Name
CATHETER, CANNULA AND TUBING, VASCULAR, CARDIO
Manufacturer (Section D)
PERFUSION SYSTEMS
7611 northland dr
brooklyn park MN 55428
Manufacturer (Section G)
PERFUSION SYSTEMS
7611 northland dr
brooklyn park MN 55428
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key17245586
Report Number2184009-2023-00743
Device Sequence Number17852521
Product Code DWF
UDI-Device Identifier00763000135690
UDI-Public00763000135690
Combination Product (Y/N)N
Initial Reporter StateNE
Initial Reporter CountryUS
PMA/510(K) Number
K150422
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Expiration Date08/05/2023
Device Model Number78220
Device Catalogue Number78220
Device Lot Number2020080264
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/15/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/06/2021
Initial Report FDA Received Date07/03/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/26/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-