• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH AVALON FM50 FETAL MONITOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH AVALON FM50 FETAL MONITOR Back to Search Results
Model Number 865071
Medical Device Problem Code Incorrect Measurement (1383)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/07/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Philips is in the process of obtaining additional information regarding the reported event and the investigation is ongoing.A follow-up report will be submitted upon completion of the investigation.
 
Event or Problem Description
The customer reported that the spo2 numbers not correct and reliabel from the device.The device was in use at time of event, there was no adverse event reported.
 
Additional Manufacturer Narrative
After further investigation this complaint is deemed no longer a reportable event.It was reported the device presented with the following: spo2 numbers not correct and reliable.The customer stated that the spo2 numbers not correct and reliable.The philips field service engineer was at the customer site for onsite service.Objective evidence was provided that demonstrates no malfunction occurred, then under such a case the device is operating as designed and as configured.It was confirmed that the reported issue was caused by an non-philips spo2 sensor.There was no report of a death or serious injury, nor was there a report of any adverse impact to any user or patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AVALON FM50 FETAL MONITOR
Common Device Name
AVALON FM50 FETAL MONITOR
Manufacturer (Section D)
PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM  71034
Manufacturer (Section G)
PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM   71034
Manufacturer Contact
hauke schik
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM   71034
70314632031
MDR Report Key17245762
Report Number9610816-2023-00325
Device Sequence Number17457828
Product Code HGM
UDI-Device Identifier00884838000438
UDI-Public00884838000438
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K140535
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number865071
Device Catalogue Number865071
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/07/2023
Supplement Date Received by Manufacturer06/07/2023
Initial Report FDA Received Date07/03/2023
Supplement Report FDA Received Date07/31/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-