Model Number 90620US
Medical Device Problem Code Overheating of Device (1437)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The duo headlight 2 bay system - us (90620us) was returned for evaluation: failure analysis: the returned duo headlight was received in used condition.Evaluation could not identify any issues with the headlight overheating or flickering; however, evaluation did identify that the fan was loud while the unit was tested.The fan has been replaced and the unit was tested and passed all the testing.Root cause analysis: the reported complaint could not be confirmed.The issue of the unit overheating and flickering could be due to a malfunctioning fan.No manufacturing, workmanship, or material deficiency has been identified.
 
Event or Problem Description
It was reported that the cooling fan of the duo headlight 2 bay system - us (90620us) operated intermittently, causing the light to overheat & flicker.It is unknown under what circumstance this event occurred; however, no patient injury, death or surgical delay was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DUO HEADLIGHT 2 BAY SYSTEM - US
Common Device Name
PFM05
Manufacturer (Section D)
INTEGRA LIFESCIENCES (IRELAND) LTD.
sragh tullamore, co. offaly
EI 
Manufacturer (Section G)
INTEGRA LIFESCIENCES (IRELAND) LTD.
sragh tullamore, co. offaly
EI  
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key17247220
Report Number3006697299-2023-00070
Device Sequence Number20510324
Product Code FSR
UDI-Device Identifier10381780490746
UDI-Public10381780490746
Combination Product (Y/N)N
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 07/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number90620US
Device Catalogue Number90620US
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/02/2023
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/12/2023
Initial Report FDA Received Date07/03/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-