ARTHREX, INC. FIBERTAK WITH SUTURETAPE; FASTENER, FIXATION, NON-DEGRADABLE, SOFT TISSUE
|
Back to Search Results |
|
| Model Number |
FIBERTAK WITH SUTURETAPE |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
12/01/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On 06/06/2023, it was reported by a sales representative via email that an ar-3602 fibertak inserter broke off in the patient's bone.This was discovered during a six-month post-operative visit.The procedure was a glenoid instability repair, and it was performed in (b)(6) 2022.No further information has been provided.Additional information requested.
|
| |
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event has not yet been determined as the device has not been returned for evaluation at this time.A follow-up report will be submitted, including a most likely cause if a root cause can not be determined.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to misuse due to applying excessive mechanical forces upon insertion, improper bone prep and/or prying/leveraging the device during use.Per dfu-0054-eo revision 5, e.Warnings: 7.Avoid excessive impaction during anchor insertion as this could lead to inserter damage and/or breakage.If insertion resistance is encountered, do not impact harder.Replace the implant and repeat the drilling/insertion process.8.Visually inspect all devices immediately after use for the potential of loose body remnants left behind in the patient.
|
| |
|