On (b)(6) 2023, it was reported by a sales representative via email that an ar-13280-35 and ar-13280-50 cancellous screw broke post-operative.This was discovered during a post-operation follow-up x-ray on (b)(6) 2023.The patient has complained of a sensation of instability but does not report a specific event or time frame in which the instability started.The implants are still in the patient, and revision surgery has not occurred yet.Additional information received on (b)(6) 2023: the original procedure occurred on (b)(6) 2022, and revision surgery occurred on (b)(6)2023.Additional information received on (b)(6)2023: during the revision surgery, an abs-3016 synthetic bone filler, an ar-13200st-17.5 tibial a/p sloped osteotomy plate, three ar-13370-4 osferion osteotomy wedge, an ar-13380-40 cortical screw, and an ar-13380-46 cortical screw were explanted along with the broken ar-13280-35 and ar-13280-50 cancellous screw.Both surgeries were performed at the same facility and by the same surgeon.
|
|
Based on the image submitted for evaluation from the field, it is evident that the screw was broken on the distal end.Consequently, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to a patient-specific event.
|