| Model Number |
E-9F |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/22/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Olympus received medwatch report mw5118174 which stated that a during bladder stone lithotripsy, the physician identified a small piece of tip from this electrohydraulic lithotripsy probe eroded.Bladder irrigation was done to remove the piece of tip.The intended procedure was completed with a similar device.The device was inspected before use and was intact.The patient was discharged home.There were no reports of patient or user harm associated with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
This device will not be returned for evaluation.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or when additional information becomes available.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 1 year since the subject device was manufactured.Based on the results of the investigation, since the device was not returned for evaluation, the reported failure could not be confirmed.Therefore, the root cause of the event could not be determined.This supplemental report includes information added to h4.Olympus will continue to monitor field performance for this device.
|
| |
|