| Model Number |
72290125 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Codes |
Perforation (2001); Rupture (2208); Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
06/08/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H10: internal complaint reference (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during a btb (bone patellar bone) acl reconstruction for revision surgery, the tip of the q-fix mini drill broke off whilst drilling a hole for let (lateral extra tenodisis) fixation.The surgeon had placed soft silk (metal) screws for femoral & tibial fixation of btb graft and it seemed that the drill tip sheared off after making contact with the screw.The bit of metal was left in the patient's femur as it could not be removed and located adjacent to the metal screw.The procedure was completed without surgical delay using a back-up device to drill an additional bone hole further down from the first hole site.No further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H3, h6: a device deficiency was not identified, and the root cause of the reported event could not be determined since the device was not returned for evaluation.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found no similar reported events.The instructions for use were reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A clinical review states that micro-motion or movement cannot be predicted and that there is potential risk for tissue inflammation and/or pain.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.
|
| |
|
Search Alerts/Recalls
|