Product analysis: the closurefast device was returned.Lot #223390147 was confirmed on the device catheter label.No ancillary devices or images were returned for review with the device.Heating element/coil condition: damage was noted along the heating coil/element.Microscopic examination revealed burnt bilogics/bubbling of the fep layer of the heating element/coil.Examination also revealed the heating element/coil of the device was seemingly exposed.The catheter cable connector was examined: the catheter reference key shows evidence of wear, and the red ink was visible in some areas.All pins were visually inspected to be straight, and no anomalies were noted along the catheter shaft.The catheter connector was plugged into the resistance tester to perform continuity/resistance and connected to rfg for functional testing; the catheter was tested according to the test parameters, test methods, and acceptance criteria.All 4 tests passed as per the acceptance criteria.The catheter was connected and recognized by both the rfg3 and rfg2 lab generators when plugged in.When the temperature rose to 120 c, the device completed the cycle without seeing an advisory message.Image analysis: one image was received of closurefast device with burnt biologics/bubbling of the fep layer apparent on the heating element/coil.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
It is reported a customer experienced an issue with a closurefast catheter.The ifu was followed during preparation, procedure, post -procedure.Local anesthesia w was used.Tumescent infiltration was utilized.One segment was treated.The coating on the catheter seemed damaged.Compression had been applied when the issue occurred.There were no alert/error message observed on the rfg but it was reported the wattage readings were somewhat abnormal on the rfg.No patient symptoms or complications associated with this event.No patient injury reported.
|