Model Number CNA0T0
Medical Device Problem Codes Malposition of Device (2616); Scratched Material (3020); Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/05/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
An other health care professional reported that lens was stuck inside the applicator, lens was also scratched.According to additional information, the lens tore and became caught in the conjunctive during insertion, and it also suffered damage after being removed from the wound.The lens was removed and replaced with another company lens in an initial procedure.There was no patient harm.
 
Additional Manufacturer Narrative
The product was returned for analysis.Additional information: the complainant indicates the use of non company viscoelastic, which is not qualified for use with company suspect device.The reported complaint was not observed as no sample was returned for analysis; however, based on the information provided by the customer, the most likely root cause is failure to follow ifu, as the surgeon states the use of non-qualified viscoelastic.The use of an unqualified combination may cause damage to the iol and potential complications during the implantation process.Based on the results from the product history record, the products met release criteria.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
The returned sample is not the actual complaint unit.Sample of a different serial number was returned.No clarification was received from the complaint.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLAREON IOL WITH THE AUTONOME DELIVERY SYSTEM
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI  00000
Manufacturer (Section G)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI   00000
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key17254614
Report Number9612169-2023-00499
Device Sequence Number5911081
Product Code HQL
UDI-Device Identifier00380652393812
UDI-Public00380652393812
Combination Product (Y/N)N
Initial Reporter CountrySF
PMA/510(K) Number
P190018
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/27/2024
Device Model NumberCNA0T0
Device Lot Number25168654
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/07/2023
Supplement Date Received by Manufacturer08/29/2023
10/31/2023
Initial Report FDA Received Date07/05/2023
Supplement Report FDA Received Date09/26/2023
11/27/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/29/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ACRYSOF IOL; AUROGEL VISCO ELASTIC (3B130)
Patient Age77 YR
Patient SexFemale
-
-