| Model Number |
20472100 |
| Medical Device Problem Code |
Overheating of Device (1437)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/01/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Estimated date of incident (b)(6) 2023.06jun2023 it was reported that charger burnt at usb plug in on back of charger.No fire but visibly burnt.Images obtained of the device showed a melted charger port.No harm to people or property reported.Original equipment was used.Device is out of warranty.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturer's ref# (b)(4).Trended case concluded: the connector is broken, probably because the user has (not intentionally) misused it, while inserting the cable.If the cable has been slightly out of specification that could have had a negative effect on this.This has caused a low ohmic connection in the cable connector - almost a short circuit - that will emit heat when the cable is inserted into the wall charger, causing up to the rated 5 w of the wall charger to be dissipated inside the connector.Micro usb is a very well known and used standard but can mechanically break down.Manufacturer's investigation is ongoing, a final report will be submitted on completion.This is an initial report.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturer's ref# (b)(4).Trended case concluded: the connector is broken, probably because the user has (not intentionally) misused it, while inserting the cable.If the cable has been slightly out of specification that could have had a negative effect on this.This has caused a low ohmic connection in the cable connector - almost a short circuit - that will emit heat when the cable is inserted into the wall charger, causing up to the rated 5 w of the wall charger to be dissipated inside the connector.Micro usb is a very well known and used standard but can mechanically break down.Manufaacturer's investigation is ongoing, a final report will be submitted on completion.This is an initial report.(b)(6)2023 technical investigation concluded: a dhr review have been completed.No deviation or changes were found during manufacturing of the device.Device investigation: confirms trended case conclusion.The clinical investigation concluded: it has been reported that the charger burnt at the usb port.There has been no harm to the user, and there is no mention of property damage.Original accessories used.The chargers have been tested according to applicable safety standards.These risks are reduced as far as possible and the probability of occurrence of these events is considered very low.In case of moisture, sweat or dirt in the connector, there can be high temperatures inside the connector due to power dissipation when connected to the charger plug.There can be melting of the plastic around the connector and is visible to the naked eye.It is highly unlikely that the user would touch in case there appears to be melting or high temperatures.In the unlikely event of the user touching this overheated device, it can lead to superficial or minor injury that would in most cases require no medical intervention.There are no new clinical aspects (e.G., unforeseen use or clinical conditions) discovered from this event.Therefore, it is concluded that the current clinical evaluation and risk/benefit conclusion herein are sufficient.Risk register conclusion reference: the residual risk after the risk control measure(s) is acceptable.No new hazardous situations revealed and no need to update the risk register.Manufacturer's investigation is completed.This is a final report.
|
| |
|
Event or Problem Description
|
|
Estimated date of incident (b)(6)2023 (b)(6)2023 it was reported that -charger burnt at usb plug in on back of charger.No fire but visibly burnt.Images obtained of the device showed a melted charger port.No harm to people or property reported original equipment was used device is out of warranty.(b)(6)2023 no further follow up received.
|
| |
|