This programmer was thoroughly inspected and analyzed upon receipt at our quality assurance laboratory.Functional testing and baseline checks were completed.The programmer was powered on and the log file was downloaded.The programmer was able to interrogate a test pulse generator device several times, confirming successful communication between the programmer and device.The ecg and pacing system analyzer (psa) functions passed all testing.Additionally, the touchscreen was tested for functionality and calibration.No anomalies were observed.There was an error recorded in the programmers log file.And benchtop testing was unable to reproduce the behavior.An unresponsive touchscreen is a known and anticipated potential hazard that can occur during use of the device, which is accounted for in device risk documentation and labeling.Boston scientific has received similar reported events where the latitude programmer screen became unresponsive to touch.And a complaint review confirmed, that the occurrence of harm associated with these events is low.Investigation of this behavior has not found any commonalities (i.E., device serial numbers, date of the event, date of manufacture, etc.) in the reports received.
|