• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON, DICKINSON & CO. (BROKEN BOW) BD VACUTAINER® URINE COMPLETE CUP KIT; TRANSPORT CULTURE MEDIUM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON & CO. (BROKEN BOW) BD VACUTAINER® URINE COMPLETE CUP KIT; TRANSPORT CULTURE MEDIUM Back to Search Results
Model Number 364956
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/13/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during use with the bd vacutainer® urine complete cup kit, samples leaked out.There was no report of patient impact.The following information was provided by the initial reporter: customer states cups are leaking frequently and they had to stop using them.The staff would collect the urine and attempt to screw the cap back on the cup and had difficult with getting the cap to completely screw on.Did the specimen(s) need to be recollected? on occasion.Did exposure to blood/ bf occur? no.
 
Additional Manufacturer Narrative
H3.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Additional Manufacturer Narrative
H6.Investigation summary: no samples or photos were returned by the customer in support of this complaint from catalog 364956, lot number is unknown.The retentions could not be inspected as the lot number is unknown.Bd was unable to duplicate or confirm the customer¿s indicated failure mode because no samples or photos were returned, and lot number is unknown.Lot number is unknown; therefore, no retention samples are available.The device history records could not be reviewed as the lot number is unknown.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business team regularly reviews the collected data for identification of emerging trends.H3 other text : see h10.
 
Event or Problem Description
It was reported that during use with the bd vacutainer® urine complete cup kit, samples leaked out.There was no report of patient impact.The following information was provided by the initial reporter: customer states cups are leaking frequently and they had to stop using them.The staff would collect the urine and attempt to screw the cap back on the cup and had difficult with getting the cap to completely screw on.Did the specimen(s) need to be recollected? on occasion.Did exposure to blood/ bf occur? no.If exposure occurred was there any post exposure testing or medical intervention? na.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® URINE COMPLETE CUP KIT
Common Device Name
TRANSPORT CULTURE MEDIUM
Manufacturer (Section D)
BECTON, DICKINSON & CO. (BROKEN BOW)
150 south 1st avenue
broken bow NE 68822
Manufacturer (Section G)
BECTON, DICKINSON & CO. (BROKEN BOW)
150 south 1st avenue
broken bow NE 68822
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8448235433
MDR Report Key17259424
Report Number1917413-2023-00631
Device Sequence Number7911658
Product Code JSM
UDI-Device Identifier50382903649568
UDI-Public50382903649568
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K024240
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number364956
Device Catalogue Number364956
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/22/2023
Supplement Date Received by Manufacturer07/07/2023
Initial Report FDA Received Date07/05/2023
Supplement Report FDA Received Date07/11/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-