| Model Number |
71938-01 |
| Medical Device Problem Code |
Product Quality Problem (1506)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/13/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The customer¿s product has been requested for investigation.A follow-up report will be filed once additional information is obtained.All pertinent information available to abbott diabetes care has been submitted.
|
| |
|
Event or Problem Description
|
|
Customer reports their adc device is too hot to hold.No further details are available at this time.It is unknown if the customer's device was too hot to hold during use or while charging.There was no report of fire, smoke, explosion, or shock.It is unknown at this time if the charging cable and adapter used was the cable and adapter supplied by adc for use with the libre device.There was no report of death, serious injury, or mistreatment associated with this event.Issues involving thermal events occurring with adc freestyle libre and libre 2 readers is associated with report of corrections and removal number 2954323-02/09/23-001-c, submitted to the fda on 09-feb-23.Adc field action fa1005-2023 was issued to the us 09feb23.
|
| |
|
Additional Manufacturer Narrative
|
|
Reader (b)(6) has been returned and investigated.Visual inspection has been performed on the returned reader and liquid contamination was observed.The returned reader was further investigated and de-cased.Visual inspection has been performed on the de-cased reader's pcba (printed circuit board assembly) and no evidence of fire was observed; however, liquid contamination was observed on the reader pcba.The returned battery voltage was measured to be within specification.Reader was also monitored under thermal camera during operation, where no abnormal hot spots were observed.Unable to perform power consumption test due to liquid contamination.The customer did not return the usb charging cable and adapter.There was no evidence observed that would cause the reader to become ¿too hot to hold" as reported by the customer.Power consumption test did not performed due to liquid contamination.Therefore, the issue is confirmed to use.An extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.The dhrs (device history review) for the libre reader and verification of correct cable and adapter were reviewed.The dhrs showed that the libre reader passed all tests prior to release and the correct cable and adapter were part of the kit pack.If partial product is returned, the case will be re-opened, and a physical investigation will be performed.All pertinent information available to abbott diabetes care has been submitted.
|
| |
|
Event or Problem Description
|
|
Customer reports their adc device is too hot to hold.No further details are available at this time.It is unknown if the customer's device was too hot to hold during use or while charging.There was no report of fire, smoke, explosion, or shock.It is unknown at this time if the charging cable and adapter used was the cable and adapter supplied by adc for use with the libre device.There was no report of death, serious injury, or mistreatment associated with this event.Issues involving thermal events occurring with adc freestyle libre and libre 2 readers is associated with report of corrections and removal number 2954323-02/09/23-001-c, submitted to the fda on 09-feb-23.Adc field action fa1005-2023 was issued to the us 09feb23.
|
| |
|
Search Alerts/Recalls
|