MONTERIS MEDICAL 3.3MM MINI-BOLT VUE Back to Search Results
Model Number 21133
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code Device Embedded In Tissue or Plaque (3165)
Date of Event 06/07/2023
Type of Reportable Event Serious Injury
Event or Problem Description
During an ablation procedure, it was reported that a 3.3mm vue mini-bolt was used.When anchoring the vue bolt, the user did a total of 5.5 turns after start of purchase.For vue bolt removal, the user used the blue hex adapter.The blue adapter could not grip and would only spin on top of the vue bolt.Instead, the user tried to use the vue bolt removal end of the removal tool after given specific instructions to prevent torque to the bolt.The removal tool was turned once, then an audible noise was heard on the second turn.The tool was removed from the bolt and the ceramic shards were visible.The bolt was anchored in the left frontal region above the hairline.The vue bolt was placed close to perpendicular to the skull with no more than a 10 degree angle.The user was standing on the right side of the patient facing superiorly reaching across to the left side from above the patient when using the vue end of the tool to remove the vue bolt.The clinical specialist confirmed that the user first attempted to use hemostats around the vue bolt to remove it from the skull, however, it kept slipping.The user then used a small drill to make a few holes around the base of the bolt around one side.Then a curette was used to remove remaining bone.The user was then able to easily remove the vue bolt from the bone without further damage to the bolt.The vue bolt, broken ceramic shards, and hex adapter was sent back to the office for investigation.
 
Additional Manufacturer Narrative
The vue bolt, fragmented pieces of the bolt, and the hex adapter was returned, and the bolt was observed to be broken on the crown.An investigation was done to investigate potential failure modes of the vue mini-bolt removal.Testing results of the investigation showed that without the application of side loading, the samples were able to withstand higher force during removal.Investigation determined that when the removal tool is not properly seated or aligned on the vue mini-bolt, the multi-bending forces and stress concentricity can cause the bolt-crown to break.There are no additional patient comorbidities reported for this case.
 
Event or Problem Description
During an ablation procedure, it was reported that a 3.3mm vue mini-bolt was used.When anchoring the vue bolt, the user did a total of 5.5 turns after start of purchase.For vue bolt removal, the user used the blue hex adapter.The blue adapter could not grip and would only spin on top of the vue bolt.Instead, the user tried to use the vue bolt removal end of the removal tool after given instructions by monteris clinical specialist to prevent torque to the bolt with the metal end of the tool.The removal tool was turned once, then a noise was heard on the second turn.The tool was removed from the bolt and the ceramic shards were visible.The bolt was anchored in the left frontal region above the hairline.The vue bolt was placed close to perpendicular to the skull with no more than a 10 degree angle.The user was standing on the right side of the patient facing superiorly reaching across to the left side from above the patient when using the vue end of the tool to remove the vue bolt.The user first attempted to use hemostats around the vue bolt to remove it from the skull, however, it kept slipping.The user then used a small drill to make a few holes around the base of the bolt around one side.Then a curette was used to remove remaining bone.The user was then able to easily remove the vue bolt from the bone without further damage to the bolt.The vue bolt, broken ceramic shards, and hex adapter was sent back to the office for investigation.
 
Additional Manufacturer Narrative
The vue bolt, fragmented pieces of the bolt, and the hex adapter was returned, and the bolt was observed to be broken on the crown.An investigation was done to investigate potential failure modes of the vue mini-bolt removal.Testing results of the investigation showed that without the application of side loading, the samples were able to withstand higher force during removal.Investigation determined that when the removal tool is not properly seated or aligned on the vue mini-bolt, the multi-bending forces and stress concentricity can cause the bolt-crown to break.Additional information from the physician was received on 6-jul-2023, confirming that there were no patient complications reported for this case, and the patient was discharged according to plan.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3.3MM MINI-BOLT VUE
Common Device Name
3.3MM MINI-BOLT VUE
Manufacturer (Section D)
MONTERIS MEDICAL
131 cheshire ln
suite 100
minnetonka MN 55305
Manufacturer Contact
jennifer englund
131 cheshire ln
suite 100
minnetonka, MN 55305
MDR Report Key17261817
Report Number3009970070-2023-00020
Device Sequence Number17656946
Product Code HAO
UDI-Device Identifier00816589021233
UDI-Public00816589021233
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
N/A
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 07/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/29/2025
Device Model Number21133
Device Catalogue NumberCMB033-V
Device Lot Number8379601
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/13/2023
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/07/2023
Supplement Date Received by Manufacturer06/07/2023
Initial Report FDA Received Date07/05/2023
Supplement Report FDA Received Date07/17/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/29/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-