• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALERE SAN DIEGO, INC. HCG CASSETTE RAPID TEST; VISUAL, PREGNANCY HCG, PRESCRIPTION USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALERE SAN DIEGO, INC. HCG CASSETTE RAPID TEST; VISUAL, PREGNANCY HCG, PRESCRIPTION USE Back to Search Results
Model Number FHC-102
Medical Device Problem Code False Negative Result (1225)
Health Effect - Clinical Code Hormonal Imbalance (4495)
Date of Event 03/09/2023
Type of Reportable Event Serious Injury
Event or Problem Description
The distributor reported an unspecified number of false negative results when testing patient samples using the hcg cassette rapid test, but was only able to provide information for one patient.The patient was tested as a screen prior to receiving a depo-provera injection for contraception.The patient tested negative on the hcg cassette rapid test and received the depo-provera injection.An unspecified urine pregnancy test was performed at another clinic 26 days later with a positive result.A quantitative blood hcg test was performed the following day with a result of 41,863 (units not specified), and the patient was diagnosed as pregnant.The patient's progesterone was determined to be low (5.8, units not specified); the patient was prescribed 200 mg progesterone per day.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HCG CASSETTE RAPID TEST
Common Device Name
VISUAL, PREGNANCY HCG, PRESCRIPTION USE
Manufacturer (Section D)
ALERE SAN DIEGO, INC.
9942 mesa rim rd
san diego 92121
MDR Report Key17263530
Report Number3006984151-2023-00002
Device Sequence Number5911383
Product Code JHI
UDI-Device Identifier20885380020251
UDI-Public(01)20885380020251(17)240531(10)HCG2062030
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Importer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 07/05/2023
Report Date (Section F) 07/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2024
Device Model NumberFHC-102
Device Catalogue NumberB1077-22
Device Lot NumberHCG2062030
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date04/11/2023
Event Location Outpatient Treatment Facility
Date Report to Manufacturer04/11/2023
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date07/05/2023
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age24 YR
Patient SexFemale
Patient Weight47 KG
Patient RaceBlack Or African American
Date Report Sent to FDA07/05/2023
-
-