OLYMPUS WINTER & IBE GMBH GRASPING FORCEPS, 5 FR., SEMI-FLEXIBLE, STANDARD Back to Search Results
Model Number A2500
Medical Device Problem Codes Break (1069); Mechanical Problem (1384)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/06/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was not returned to olympus for evaluation, however, photo of the subject device was provided for evaluation.In addition, according to the report via phone, the field service engineer (fse) physical inspection found the device had broken jaws , broken connection to the control rod.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over a year since the subject device was manufactured.Based on the results of the investigation, it is likely that the following led to the malfunction: improper handling by the user.Olympus will continue to monitor the field performance of this device.
 
Event or Problem Description
The customer reported to olympus, the grasping forceps had a broken end.The issue was found during reprocessing by a nurse.There were no reports of patient harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GRASPING FORCEPS, 5 FR., SEMI-FLEXIBLE, STANDARD
Common Device Name
FORCEPS
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, deutschland 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, deutschland 22045
GM   22045
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17264326
Report Number9610773-2023-01806
Device Sequence Number17613686
Product Code GCP
UDI-Device Identifier04042761003483
UDI-Public04042761003483
Combination Product (Y/N)N
Initial Reporter CountryEZ
PMA/510(K) Number
K790071
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 07/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberA2500
Device Catalogue NumberA2500
Device Lot Number21X01
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/07/2023
Initial Report FDA Received Date07/06/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/03/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-