• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER NEUROVASCULAR-UTAH-SALT LAKE CITY SYNCHRO2-14 STANDARD STRAIGHT 300CM; WIRE, GUIDE, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER NEUROVASCULAR-UTAH-SALT LAKE CITY SYNCHRO2-14 STANDARD STRAIGHT 300CM; WIRE, GUIDE, CATHETER Back to Search Results
Model Number M00326510
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/13/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a neurovascular procedure the subject guidewire could not pass the ica (internal carotid artery) siphon section to distal occlusion.The subject guidewire was retrieved and it was found that fractured at tip with core wire connected.70 percentage stenosis and calcification at the lesion was reported.The subject guidewire was replaced, and the procedure was completed successfully.No clinical consequences were reported to the patient due to this event.
 
Additional Manufacturer Narrative
This is 2 of 2 reports (2nd mdr) h3 other text : the device is not available to the manufacturer.
 
Event or Problem Description
It was reported that during a neurovascular procedure the subject guidewire could not pass the ica (internal carotid artery) siphon section to distal occlusion.The subject guidewire was retrieved and it was found that fractured at tip with core wire connected.70 percentage stenosis and calcification at the lesion was reported.The subject guidewire was replaced, and the procedure was completed successfully.No clinical consequences were reported to the patient due to this event.
 
Additional Manufacturer Narrative
There are controls in the manufacturing process to ensure the product met specifications upon release.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The event was confirmed based on the damage noted to the returned device.The device failed to meet specification when returned for analysis.Analysis of the returned device found the guidewire niti tubing was fractured with stretching noted to the platinum coil.The damage noted to the returned device is indicative of excessive forces being applied to the guidewire, most likely during the attempt to retract the device from the patients anatomy.Additional information provided by the customer indicated the device was prepared for use as per the directions for use, the device was confirmed to be in good condition during preparation/prior to use on the patient, continuous flush was set up and maintained throughout the clinical procedure and the patients anatomy was not tortuous.There was no resistance encountered during the usage of the device that could have contributed to the issue, there was 70% stenosis and calcification at the lesion.An assignable cause of procedural factors will be assigned to the as reported ¿guidewire distal tip broken/fractured during use¿ and to the analyzed 'guidewire distal tip stretched' and 'guidewire broken/fractured during use' as the issue is associated with a product that meets stryker design and manufacture specifications and was used in according with the dfu but due to procedural and/or anatomical factors during use, the product performance was limited.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCHRO2-14 STANDARD STRAIGHT 300CM
Common Device Name
WIRE, GUIDE, CATHETER
Manufacturer (Section D)
STRYKER NEUROVASCULAR-UTAH-SALT LAKE CITY
4870 west 2100 south
salt lake city UT 84120
Manufacturer (Section G)
STRYKER NEUROVASCULAR-UTAH-SALT LAKE CITY
4870 west 2100 south
salt lake city UT 84120
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key17264476
Report Number3012931345-2023-00153
Device Sequence Number15424430
Product Code DQX
UDI-Device Identifier04546540688781
UDI-Public04546540688781
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K053268
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/17/2025
Device Model NumberM00326510
Device Catalogue NumberM00326510
Device Lot Number0000209222
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/13/2023
Supplement Date Received by Manufacturer08/15/2023
Initial Report FDA Received Date07/06/2023
Supplement Report FDA Received Date09/12/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/19/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age45 YR
Patient SexMale
Patient Weight50 KG
Patient RaceAsian
-
-