| Model Number |
03P6822 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/14/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer observed falsely elevated magnesium results generated on the architect c4000 processing module for one sample that was not released out of the lab.The following results were provided: sid (b)(6) original result = 7.5 mg/dl, rerun result on another architect = 1.9 mg/dl no impact to patient management was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.Completed information for section a1 patient identification: sids(b)(6) all available patient information was included.Additional patient details are not available.
|
| |
|
Event or Problem Description
|
|
The customer observed falsely elevated magnesium results generated on the architect c4000 processing module for one sample that was not released out of the lab.The following results were provided: sid (b)(6)original result = 7.5 mg/dl, rerun result on another architect = 1.9 mg/dl no impact to patient management was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The complaint investigation for false elevated magnesium results included a review of data and information provided by the customer, search for similar complaints, ticket trending review, labeling review, and device history record review.Return testing was not completed as returns were not available.Instrument troubleshooting including cleaning the cuvettes, rebuilding reagent and sample syringes, and other adjustments/replacements were performed.Trending review did not identify any trends for the issue for the product.The device history record review did not identify any non-conformances or deviations with lot number 27365un22 and the complaint issue.Labeling was reviewed and sufficiently addresses the customer's issue.Based on the investigation, no systemic issue or deficiency with the magnesium assay for lot 27365un22 was identified.All available patient information was included.Additional patient details are not available.
|
| |
|